Granite Temp 120

20-Channel Thermocouple Module
Granite Temp 120

Granite Temp 120

OrderableConfigurableGSA

20-Channel Thermocouple Module

Configuration

Sign in for option pricing

Select a configuration to add to your quote

Model:
Granite Temp 120
Item Number:
Temp120
Warranty:
1 year
Lead time:
45 days

Overview

GRANITE Series 20-Channel Thermocouple Module

The Granite™ Temp 120 is a purpose-built thermocouple measurement module. It has 20 thermocouple inputs using universal-alloy miniature thermocouple jacks. The Temp 120 is made to handle extreme temperature environments and rapid thermal gradients applied to the measurement module. It has a dedicated temperature reference for every input terminal and is carefully designed to reduce temperature errors in the presence of external temperature gradients.

Key Advantages

Better Temperature Measurements
With dedicated reference temperatures

Technical Description

The inputs of the Temp 120 are a universal copper-copper junction and can accept any standard thermocouple alloy type. The Temp 120 is part of a data acquisition system that includes a DAQ such as a Granite 6, 9, or 10, or the CR6 or CR1000X Dataloggers. The clock timing synchronization across modules is managed by the DAQ or data logger via the CPI communication port. The time base can be referenced either to GPS or a temperature-compensated real-time clock. Applications requiring hundreds of terminals of thermocouples can be easily assembled as a single data acquisition system. Measurement modules can be collocated or spread out via a networking cable to be closer to the point of measurement.

Benefits

  • Universal-type standard miniature thermocouple connectors
  • Cold-junction reference temperatures on every channel
  • Exceptional accuracy and temperature stability
  • CPI DAQ connection for channel expansion
  • CAN bus 2.0 A/B output available
  • USB 2.0 interface for PC-based operation

Specifications

Operating Temperature Range
-40° to +70°C

(standard)

Operating Temperature Range
-55° to +85°C

(extended)

Terminals
20 channels: universal (copper/copper), miniature socket connectors
Supported Thermocouple Types
B, E, J, K, N, R, S, T
Resolution
0.015°C
Speed
10 Hz when filtering is disabled

(maximum measurement rate for all 20 channels)

Speed
1 Hz when filtering is enabled

(maximum measurement rate for all 20 channels)

Input Limits
0.25 to 3.0 Vdc

(built-in 1.65 Vdc bias voltage)

Maximum Input Voltage
±10 Vdc
DC Common-Mode Rejection
≥ 115 dB
Normal Mode Rejection
> 80 dB

(at 50 and 60 Hz)

Internal Cold Junction Accuracy
±0.05°C typical

(20° to 40°C)

Internal Cold Junction Accuracy
±0.28°C maximum

(-40° to +70°C) for -SD option

Internal Cold Junction Accuracy
±0.36°C maximum

(-55° to +85°C) for -XD option

Total Measurement Accuracy in Static Temperature Environment
±0.4°C typical

(20° to 40°C)

Total Measurement Accuracy in Static Temperature Environment
±0.8°C maximum

(-40° to +70°C) for -SD option

Total Measurement Accuracy in Static Temperature Environment
±1.0°C maximum

(-55° to +85°C) for -XD option

Total Measurement Accuracy for Rapid Temperature Environment Ramps
±3°C maximum

(8°C per minute ambient gradient)

Dimensions
21.5 x 10.8 x 5.1 cm

(8.5 x 4.25 x 2 in.)

Weight
0.9 kg

(1.95 lb)

Measurement Performance Verification and Calibration

-SD Option
Performance is based on engineering design and verified in a laboratory environment.
-XD Option
Performance is based on engineering design and verified over the full temperature range of the product and includes a Certificate of Calibration and Instrument Data Report.

Communications

CPI
RJ45 interface to Campbell Scientific data loggers and CDM measurement peripherals and sensors
USB
USB micro-B device only, 2.0 full-speed 12 Mbps, for computer connection

System

Processor
Renesas RX63N

(32-bit with hardware FPU, running at 96 MHz)

Memory
2 MB SRAM
A/D Converter
24-bit sigma-delta

Power Requirements

Voltage
9.6 to 32 Vdc
Typical Current Drain
5 mA

(sleep)

Typical Current Drain
30 mA

(active)

Ships With

(2)

Related Resources

Connect With Us

envelope icon
Phone Numbers
General: (435) 227-9000
location icon
Corporate Address
815 W 1800 N
Logan, UT 84321
send icon
Email Address
sales@campbellsci.com

As a trusted leader in measurement systems for more than 50 years, our top priority at Campbell Scientific has always been to provide the most accurate, reliable data for research and industry.