SI-111SS

Infrared Radiometer with Standard Field of View
SI-111SS

SI-111SS

OrderableConfigurable

Infrared Radiometer with Standard Field of View

Best for general non-contact applications.

Model:
SI-111SS
Item Number:
34452
Warranty:
4 years
Lead time:
14 days

Overview

Infrared Radiometer with Standard Field of View

The SI-111SS, manufactured by Apogee, is a precision infrared radiometer that determines the surface temperature of an object without physical contact. It measures both the subject's surface temperature and the sensor-body temperature. A Campbell Scientific data logger uses these measurements to calculate the correct temperature of the subject.

This radiometer features an IP67-rated, marine-grade 316L connector that allows the user to easily swap sensors for recalibration or to replace damaged cables. 

Key Advantages

Determine an Object's Surface Temperature
No physical contact necessary

Technical Description

The SI-111SS consists of a thermopile, which measures surface temperature, and a thermistor, which measures sensor body temperature. The two temperature sensors are housed in a rugged aluminum body that contains a germanium window.

Both the thermopile and the thermistor output a millivolt signal that most of our data loggers can measure. The data logger uses the Stefan-Boltzman equation to correct for the effect of sensor body temperature on the target temperature. The corrected readings yield an absolute accuracy of ±0.2°C from -10° to +65°C.

Field of View (FOV)

The SI-111SS has a 22-degree half-angle field-of-view (FOV). The FOV is reported as the half-angle of the apex of the cone formed by the target (cone base) and the detector (cone apex). The target is a circle from which 98% of the radiation viewed by the detector is being emitted.

Benefits

  • Compatible with most Campbell Scientific data loggers
  • Measures surface temperature continuously in the field
  • Provides road surface, plant canopy, soil surface, snow surface, and water surface temperature measurements
  • Avoids influencing the temperature, providing more accurate measurements
  • Ideal for providing spatial averages
  • Rugged construction—two temperature probes housed in an aluminum body with a germanium window

Specifications

Sensor
Thermopile and thermistor
Input Power
2.5 V excitation

(for thermistor)

Response Time
< 1 s

(to changes in target temperature)

Target Temperature Output Signal
60 μV per °C difference from sensor body
Body Temperature Output Signal
0 to 2500 mV
Optics
Germanium lens
Wavelength Range
8 to 14 μm

(corresponds to atmospheric window)

Field of View (FOV)
22° half angle
Operating Temperature Range
-55° to +80°C
Operating Relative Humidity Range
0 to 100% RH
Cable Description
4.5 m (14.76 ft) twisted, shielded 4-conductor wire with Santoprene casing, ending in pigtails
Absolute Accuracy
±0.2°C

(-10° to +65°C)

Absolute Accuracy
±0.5°C

(-40° to +70°C)

Uniformity
±0.1°C

(-10° to +65°C)

Uniformity
±0.3°C

(-40° to +70°C)

Repeatability
±0.05°C

(-10° to +65°C)

Repeatability
±0.1°C

(-40° to +70°C)

Diameter
2.3 cm

(0.9 in.)

Length
6 cm

(2.4 in.)

Weight
190 g

(6.7 oz)