MS-80SH

Class A Spectrally Flat and Fast-Response Pyranometer with Integrated Dome Heater
MS-80SH

MS-80SH

OrderableConfigurable

Class A Spectrally Flat and Fast-Response Pyranometer with Integrated Dome Heater

Model:
MS-80SH
Item Number:
45080
Warranty:
5 years
Lead time:
21 days

Overview

Class A Pyranometer Spectrally Flat & Fast Response for GHI, POA, RPOA, DHI & ALBEDO (-40 to +80C) w/o Cable

The MS-80SH, manufactured by EKO Instruments, is an innovative, next-generation ISO 9060 Class A spectrally flat and fast-response (secondary standard) pyranometer. The pyranometer features a compact design with internal desiccation, a single dome over a quartz diffusor, a thermally isolated thermopile detector, negligible thermal offsets, ultra-low temperature dependency, and exceptional non-linearity characteristics. The MS-80SH has a five-year warranty and recalibration interval.

EKO instruments is the longest-operating ISO 17025-accredited pyranometer manufacturer in the world. This allows for the highest-quality calibration and for Campbell Scientific to comply with international standards (ISO/IEC 17025 for ISO 9847).

Benefits

  • ISO 9060 Class A spectrally flat with fast response (secondary standard)
  • Integrated dome heater for dew and frost mitigation as per IEC 61724-1:2021 Class A monitoring requirements
  • Ability to be combined with MV-01 external heater and ventilator in more harsh conditions
  • Industry-leading stability
  • ISO 17025-certified calibration
  • Five-year warranty and recalibration interval

Specifications

Sensor
Internal desiccation, single-dome, isolated thermopile detector, quartz diffusor
Measurement Description
Monitors solar radiation for the full solar spectrum range
ISO Classification
ISO 9060 Class A spectrally flat and fast-response pyranometer

(secondary standard)

ISO Classification
ISO 17025 Class A pyranometer
Output
Modbus RTU over RS-485

(digital)

Sensitivity
~10 µV/W/m2
Response Time
< 1 s

(95%)

Zero Offset A
< 1 W/m2

(response to 200 W/m2 net thermal radiation)

Zero Offset B
±1 W/m2

(response to 5 K/h change in ambient temperature)

Non-Stability
±0.5% change per 5 years
Non-Linearity
±0.2%

(at 1000 W/m2)

Directional Response
±10 W/m2

(at 1000 W/m2)

Spectral Selectivity
±3%

(0.35 to 1.5 μm)

Temperature Response
< 0.8%

(-10° to +40°C)

Temperature Response
< 1%

(-20° to +50°C)

Tilt Response
< ±0.2%

(0 to 90° at 1000 W/m2)

Operating Temperature Range
-40° to +80°C
Irradiance Range
0 to 4000 W/m2
Spectral Range
285 to 3000 nm
Ingress Protection
IP67