CRS456
CRS456
OrderableConfigurableTitanium Stand-Alone Pressure Transducer
Titanium recording pressure transducer for more corrosive environments.
Pricing
Configuration
Sign in for option pricing- 0-145psig (1000kPa) No Calibration Cert –SN w/Standard Nose ConeCRS456-SA-145-NC-SN
- 0-145psig (1000kPa) w/Calibration Cert –SN w/Standard Nose ConeCRS456-SA-145-CC-SN
- 0-14.5psig (100kPa) No Calibration Cert –SN w/Standard Nose ConeCRS456-SA-14-NC-SN
- 0-14.5psig (100kPa) w/Calibration Cert –SN w/Standard Nose ConeCRS456-SA-14-CC-SN
- 0-29psig (200kPa) No Calibration Cert –SN w/Standard Nose ConeCRS456-SA-29-NC-SN
- 0-29psig (200kPa) w/Calibration Cert –SN w/Standard Nose ConeCRS456-SA-29-CC-SN
- 0-72.5psig (500kPa) No Calibration Cert –SN w/Standard Nose ConeCRS456-SA-72-NC-SN
- 0-72.5psig (500kPa) w/Calibration Cert –SN w/Standard Nose ConeCRS456-SA-72-CC-SN
- 0-7.25psig (50kPa) No Calibration Cert –SN w/Standard Nose ConeCRS456-SA-7-NC-SN
- 0-7.25psig (50kPa) w/Calibration Cert –SN w/Standard Nose ConeCRS456-SA-7-CC-SN
Select a configuration to add to your quote
- Model:
- CRS456
- Item Number:
- 45301
- Warranty:
- 1 year
- Lead time:
- 14 days
Overview
The CRS456 consists of a water-level and water-temperature sensor that has its own time clock and memory to store the collected data, in a compact titanium case. This frees users to place the sensor in remote sites and let it collect data for long periods. The sensor can then be retrieved, connected to a PC via a micro-USB port, and the data transferred to the PC. HydroSci software is included and elegantly supports test setup, data retrieval, and data display. Long battery life and rugged construction mean you can trust the CRS456 to collect important data. Low cost and ease of use make it a good choice in a variety of applications. The CRS451 is the same as this, but with a stainless-steel case.
Key Advantages
Easy to use software, reliable hardware
Technical Description
The CRS456 has several pressure range options. The CRS456 is unvented; therefore it is monitoring sealed gage pressure. For maximum accuracy, the data should be corrected for barometric pressure.
HydroSci software is available for download. This software simplifies the process of configuring the CRS456. Users can configure the CRS456 to monitor surface water, ground water, or a standard pump test.
HydroSci software will also compensate the measurements for barometric pressure using an additional CS456 or a CS451 as a reference sensor, or using a barometric pressure sensor from a nearby weather station.
HydroSci software will display the data in tabular or graphical formats.
Benefits
- Sensors and data-collection features in one instrument case
- Rugged titanium case protects piezoresistive sensor allowing it to be used in saltwater and other harsh environments
- Quality construction ensures product reliability
- Fully temperature-compensated
- Fast scan rate
- Large data-storage capacity
- Long battery life
- Easy-to-use software
Specifications
- Venting
- Unvented
- Measurement Time
- < 1.0 s
- Output Options
- micro USB
- Internal Data Collection Memory
- 4 MB
- Logging/Scanning Modes
- Standard, Delta, Wave, Logarithmic
- Resolution
- 0.0035% FS
- Dry Storage Temperature Range
- -10° to +80°C
WARNING: Sensor could be damaged if encased in frozen ice. - Operating Temperature Range
- 0° to 60°C
WARNING: Sensor could be damaged if encased in frozen ice. - Overpressure
- 2 x pressure range
- Power Requirements
- Internal user-replaceable lithium battery
- Battery Life
- 5+ years
(when logging interval is once per hour)
- Body Material
- Titanium
- Distance
- 2.54 cm
(1 in.)
Distance from black line etched on housing to end of NPT fitting - Distance
- 2.3 cm
(0.9 in.)
Distance from black line etched on housing to end of standard nose cone - Distance
- 9.9 cm
(3.9 in.)
Distance from black line etched on housing to end of weighted nose cone - Water-Level Accuracy
- ±0.1% full-scale-range TEB
Includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range, per ISA S51.1. - Temperature Accuracy
- ±0.2°C
- Ingress Protection
- Exceeds IP68
- Diameter
- 2.22 cm
(0.875 in.)
- Length
- 22.23 cm
(8.75 in.)
- Weight
- 230 g
(0.51 lb)
Power Consumption
- Quiescent
- < 80 µA
- Measurement/Communication
- 4 mA
(1 s measurement)
Measurement Ranges at Fresh Water Depths
- 0 to 5.1 m (16.7 ft)
- 0 to 7.25 psi
- 0 to 5.1 m (16.7 ft)
- 0 to 50 kPa
- 0 to 10.2 m (33.4 ft)
- 0 to 14.5 psi
- 0 to 10.2 m (33.4 ft)
- 0 to 100 kPa
- 0 to 20.4 m (67 ft)
- 0 to 29 psi
- 0 to 20.4 m (67 ft)
- 0 to 200 kPa
- 0 to 50.9 m (167 ft)
- 0 to 72.5 psi
- 0 to 50.9 m (167 ft)
- 0 to 500 kPa
- 0 to 102 m (334.5 ft)
- 0 to 145 psi
- 0 to 102 m (334.5 ft)
- 0 to 1000 kPa