SnowVue 10 OS

v1.24

Current SnowVue 10 operating system. 

Note: OS update must be performed using the aXLOADOS! SDI-12 command. Please refer to the SnowVue user manual for details on the update process. 

Revision History

v1.24April 3, 2026
  • New SDI-12 Commands
    • M5/C5 - Single-ping distance.
      • Returns distance to surface using a single ultrasonic ping rather than a multi-ping median. Useful for rapid measurements or diagnostics.
      • Response: 1 value (distance in meters).
    • M6/C6 - Distance statistics.
      • Returns the median distance along with minimum, maximum, and standard deviation of the temperature-compensated distance measurements from the multi-ping cycle.
      • Response: 4 values (distance, min, max, std in meters).
    • IM5/IC5 and IM6/IC6 identify - measurement metadata responses added for the new commands.
  • SDI-12 Communication
    • SDI-12 Interference detection.
    • Improved break detection.
  • Measurement Accuracy Improvements
    • Trackback algorithm for time-of-flight measurement.
    • Updated calibration formula.
    • Tuned peak hold window.
    • Improved peak detection calculation.
    • Glitch filter.
    • Skipped pings handled gracefully.
v1.22March 13, 2025
  • Determines air/ground interface distance with better accuracy.
  • Returns an error value when a good measurement can not be made.
  • Identifies and ignores poor/noisy return echo waveforms.
v1.19February 8, 2024
  • Added XRESET command to reset sensor.
  • Reduced M/C command time to 3 seconds from 20.
  • Reduced erroneous measurements which appeared under certain conditions.
v1.18February 15, 2023
  • Updated SDI-12 responses and includes additional XHELP output.
  • Updated calibration of distance to ground.
  • Improved performance of filter used to determine distance to ground.
v1.15October 19, 2022
  • Fixed XRT! response, no longer returns -1000 when no reference temperature has been set.
  • Fixed M9! response, first value now returns reference temperature correctly.
  • Fixed resonant frequency measurement.
v1.14March 1, 2022
  • Major
    Fixed SnowVUE communication lockup issue.
  • Note: OS update must be performed using the aXLOADOS! SDI-12 command. Please refer to the SnowVUE user manual for details on the update process.
  • Minor
    Updated the SDI-12 aV! command response to include the firmware version, hardware version, and uptime.
  • Minor
    Updated the SDI-12 aI! command response to fit the new Campbell Scientific identification information.
  • Note: OS update must be performed using the aXLOADOS! SDI-12 command. Please refer to the SnowVUE user manual for details on the update process.