Current LevelVue B10 Operating System. The Device Configuration Utility is used to send this Operating System to the LevelVue B10.

Revision History

v2.3May 1, 2026
  • Added retries on some measurements when they fail.
  • Added toggling chip select around ADC accesses.
  • Added a compressor soft-start option to address units in the field that were running into a hardware current limit on compressor startup.
v2.2April 24, 2025
  • Fixed an issue with a 50PSI bubbler at a similar depth, where the pump wasn't activating, causing water to flow up the LevelVue line.
  • Fixed an issue when using a 50PSI bubbler at a similar depth, where the pump wasn't activating, allowing water to flow up the line of the LevelVue.
v2.1November 3, 2021
  • Major
    Fixed a condition where the compressor could stay in the ON state. If this were to happen, power cycling the system would correct the issue.
  • Major
    Fixed a condition where the compressor could stay in the ON state. Power cycling the system would correct the issue.