DustVue 10 OS
v1.03Current operating system for the DustVue 10.
Revision History
v1.03May 7, 2026
- Don't do cover operations below -20°C.
- Increase motor hold current.
- Change the auto threshold to 40% (from 70%) of the measured clean cell irradiance during an offsets update.
- Preserve the soiling ratios calculated during the day so they are not lost due to a power loss.
- Change daily soiling ratio calculation method - average all "Last Soiling Ratio" calculations.
- Adjust heater turn on point.
- Don't clear Sample Count register until midnight to ensure there is time to see last Sample Count increment.
- When closing the cover, move it in further to address bounce/jam issue.
- Change bit 0 in System Status Code to stay set until another cover operation is attempted and rename it "Cover Operation Error".