44301
44301
OrderableConfigurableGSAScrub Module for CPEC200/310
Pricing
- Model:
- 44301
- Item Number:
- 44301
- Warranty:
- 1 year
- Lead time:
- 30 days
Overview
The 44301 Scrub Module for the CPEC200/310 is an accessory to the CPEC200/310 Closed-Path Eddy-Covariance System. It is used to provide a source of zero air (air that contains no CO2 or H2O) to the CPEC200/310 for zeroing the EC155 gas analyzer. More often, a compressed-gas cylinder of zero air is used for this purpose; however, the CPEC200/310 scrub module may be preferable in locations where acquiring and replacing gas cylinders is not convenient.
The CPEC200/310 scrub module consists of a pump and a three-stage molecular sieve. The pump is powered via a cable (included) that connects to the CPEC200/310 system enclosure. The first stage molecular sieve should be replaced once a year to ensure the zero air is free of CO2 and H2O.
Technical Description
The CPEC200/310 Scrub Module provides a source of zero air to the CPEC200/310. Ambient air enters through an inlet at the base of the module, flows through a first stage molecular sieve, goes through a small pump, and flows through two more stages of molecular sieve. Zero air then emerges from the exhaust or outlet port.
The outlet port should be connected to the CPEC200/310 System Enclosure via tubing with 0.25 in. Swagelok® connections (pn 21823-L). The molecular sieve in the first stage cylinder should be replaced once a year to ensure the quality of zero air. Molecular sieve 13X with 1.6-2.5 mm bead diameter is recommended. Approximately 400 g (dry basis) is needed to refill one cylinder.
Specifications
- Volume
- 480 cm3
(29.29 in.3) for each molecular sieve cylinder
- Power
- 2 W
(typical)
- Dimensions
- 39.4 x 34.3 x 20.3 cm
(15.5 x 13.5 x 8.0 in.)
- Weight
- 9.6 kg
(21.1 lb)