40718
40718
OrderableCalibration Lamp for FluoreSens 10
Overview
The 40718, manufactured by Ocean Insight, is a tungsten-halogen calibrated light source covering Vis-NIR wavelengths in the 350 to 1100 nm range. The calibration lamp is radiometrically calibrated for use with a FluoreSens™10 fiber-optic cable and CC-3 cosine corrector. The 40718 provides known absolute intensity values expressed in µW/cm²/nm and in radiant flux in µW/nm.
The spectral intensity of the 40718 conforms to the National Institute of Standards and Technology (NIST) standards, meeting the calibration requirements of the FluoreSens 10 Sun-Induced Chlorophyll Fluorescence (SIF) System. The 40718 is required to calibrate the FluoreSens 10 before the system is used for the first time or anytime the fiber-optic cable is removed and reattached.
Key Advantages
Specifications
- Wavelength Range
- 350 nm to 1.1 μm
- Nominal Bulb Power
- 5 W
- Calibrated for
- Absolute irradiance
(µW/cm²/nm)
- Calibration Lifetime
- 50 hr
- Drift (Voltage)
- <0.3% per hr
- Minimum Uncertainty Level
- 3%
- Stability (Current)
- 0.5% peak to peak
- Warm-Up Time
- 15 min
- Safety and Regulatory
- CE
- Safety and Regulatory
- ROHS
- Safety and Regulatory
- WEEE
- Operating Humidity Range
- 5 to 95%
(without condensation at 40°C)
- Operating Temperature Range
- 5° to 35°C
- Shutter
- Yes
- Fiber Connector
- SMA 905
- Fiber Connector
- CC-3
- Power Consumption @ 12 Vdc
- 1.2 A
- Power Requirements
- 85 to 264 V
- Power Requirements
- 50/60 Hz
- Dimensions
- 14 x 6 x 6 cm
(5.51 x 2.36 x 2.36 in)
- Weight
- 0.5 kg
(1.1 lb)