39310

Replacement Temperature/RH Element and Cap for HygroVue 10
39310

39310

Orderable

Replacement Temperature/RH Element and Cap for HygroVue 10

Model:
39310
Item Number:
39310
Warranty:
No warranty
Lead time:
21 days

Overview

Replacement AtmosVue-HV/HygroVue10 Element & Cap

The 39310 replaces the sensor element, filter, and cap of a HygroVue™10 Digital Temperature and Relative Humidity Sensor. Replacement can take place in the field—eliminating the downtime typically required for the recalibration process. The replacement ships with a calibration verification certificate and meets the specifications of the HygroVue 10.

The 39310 is plugged into a socket underneath the end cap and filter assembly. Chip replacement may be needed every one to three years, depending on the application and environmental conditions.

Benefits

  • Replacement chip reduces downtime and maintenance costs associated with sensor calibration
  • Replacement screen restores sensor to factory performance and airflow
  • Temperature and RH (relative humidity) elements traceable to NPL and NIST standards
  • Ships with calibration verification certificate

Specifications

Sensing Element
Sensirion SHT35 enhanced by Campbell Scientific
Sensor Protection
Outer glass-filled polypropylene cap fitted with a stainless-steel mesh dust filter with nominal pore size of < 30 µm.

The sensor element has a PTFE protective film with a filtration efficiency of > 99.99% for particles of 200 nm or larger size.

Weight
22.68 g

(0.05 lb)

Relative Humidity

Measurement Range
0 to 100% RH
Accuracy
-NOTE- The accuracy figures quoted are the 95% confidence limits relative to factory standards.
Accuracy
±1.5%

(at 25°C, over the range 0 to 80% RH)

Accuracy
±2%

(at 25°C, over the range 80 to 100% RH)

Short-Term Hysteresis
< ±1% RH
Additional Errors at Other Temperatures
< ±1% RH

(over -40° to +60°C)

Long-Term Stability
±0.5% per year

(maximum drift in clean air conditions)

Reported Resolution
0.001% RH
Repeatability
0.05% RH

(3σ noise level)

Response Time with Filter
8 s

(63% response time in air moving at 1 m/s)

Air Temperature

Measurement Range
-40°C to +70°C
-NOTE-
The accuracy figures quoted are the 95% confidence limits relative to factory standards.
Accuracy
±0.2°C

(over the range -40 to +70°C)

Accuracy
±0.1°C

(over the range -20 to +60°C)

Long-Term Drift
< 0.03°C per year
Reported Resolution
0.001°C
Repeatability
0.04°C

(3σ noise level)

Response Time with Filter
< 130 s

(63% response time in air moving at 1 m/s)

Calibration Traceability
NIST and NPL standards

Maximum Current Drain

Quiescent
50 µA
During Measurement
0.6 mA

(takes 0.5 s)