OBS500
OBS500
End of lifeConfigurableSmart Turbidity Meter with Antifouling Features
Pricing
- Model:
- OBS500
- Item Number:
- 27784
- Warranty:
- 1 year
- Lead time:
- 14 days
Overview
The OBS500 is a submersible turbidity probe with active antifouling capabilities for better measurements in biologically active water with both high and low turbidity. It outputs an SDI-12, digitally processed signal that many of our data loggers can measure. ClearSensor® and OBS® are registered trademarks of Campbell Scientific.
Key Advantages
ClearSensor antifouling features, dual backscatter sensors
Technical Description
Design features of the OBS500 include the combination of a backscatter sensor (better at measuring higher turbidity) with a second sidescatter sensor (better at measuring lower turbidity). It has a shutter that is opened only during measurements, which reduces the time that algae or other organisms can cling to its optics.
To prevent biofouling and ensure better measurements, the OBS500 incorporates the ClearSensor Method (U.S. Patent No. 8,429,952). This method uses a shutter/wiper mechanism to protect and clean the optics. With the ClearSensor method, a chamber is also filled with a biocide that continuously leaches out over the optics while the probe shutter is in the closed position.
Campbell Scientific offers a disposable, plastic sleeve that can make cleanup a snap, as well as a copper sleeve that can provide additional protection, especially in sea water.
(OBS® is a registered trademark of Campbell Scientific.)
Benefits
- Dual backscatter and sidescatter sensors used to measure turbidity
- Patented ClearSensor antifouling method for better measurements in biologically active water
- Shutter/wiper mechanism keeps lenses clean
- Refillable biocide chamber prevents fouling
- Disposable plastic sleeve facilitates cleanup
- Optional copper sleeve for additional protection (especially for sea water) or disposable plastic sleeve facilitates easy cleanup
Specifications
- Dual Probe
- 90° sidescatter and backscatter
- Range
- 0 to 4000 NTU
- Active and Passive Antifouling
- Shutter, wiper, biocide, copper, optional removable sleeve
- Concentration Accuracy
- ±2% of reading or 0.5 NTU
(whichever is greater)
- Operating Temperature Range
- 0° to 40°C
- Temperature Accuracy
- ±0.3°C
- Emitter Wavelength
- 850 nm
- Power Requirements
- 9.6 to 18 Vdc
- Measurement Time
- < 10 s
- Maximum Submersion Depth
- 100 m
(330 ft)
- Diameter
- 4.8 cm
(1.88 in.)
- Maximum Cable Length
- > 500 m
(1640 ft)
- Length
- 27 cm
(10.63 in.)
- Weight
- 0.59 kg
(1.3 lb)
Power Consumption
- Quiescent
- < 200 µA
- Measurement
- < 40 mA
- Communication
- < 40 mA
- Maximum Peak Current
- 200 mA for 50 ms when shutter motor starts
- Active Shutter Motor
- < 120 mA
Outputs
- SDI-12
- Version 1.3, 1200 bps
- RS-232
- 9600 bps
- Analog
- 0 to 5 Vdc
Accessories
(10)27803
OBS500 Copper Sleeve
End of life27473
OBS500 Plastic Sleeve
End of life27225
OBS500 or OBS501 Carrying Case (holds 2)
End of life27573
Test Cable for OBS500 or OBS501
End of life27785
OBS500 or OBS501 Field Cable, 5 m (16 ft)
Orderable27786
OBS500 or OBS501 Field Cable, 10 m (32.8 ft)
Orderable27787
OBS500 or OBS501 Field Cable, 20 m (65.6 ft)
Orderable27788
OBS500 or OBS501 Field Cable, 30 m (98.4 ft)
OrderableOBS500CBL-L
OBS501 or OBS500 Field Cable
End of life20915
OBS Sensor 5-Point Sedimentation Calibration
End of life
Related Resources
Documents
- Technical Paper
OBS: Comparison of Suspended Solids Concentration (SSC) and Turbidity
April 2008
View Document - Technical Paper
OBS Measurements: Effects of Water Color
April 2008
View Document - Technical Paper
OBS: Effects of Light Absorption and Scattering in Water Samples on Measurements
April 2008
View Document - Technical Paper
OBS Sensors: Light Sources and Photodetectors
April 2008
View Document - View all documents